Diabetes technology

Post-Market Surveillance Assessment of the Clinical Accuracy of a Blood Glucose Monitoring System with an Improved Algorithm for Enhanced Product Performance

J Diabetes Sci TechnolPublished Post-market surveillance accuracy study

Authors' conclusion

This post-market surveillance program demonstrates the new BGM system consistently meets measures of clinical accuracy specified by regulators. This program supports a growing demand by regulators for real-world evidence demonstrating consistent in-market product efficacy as opposed to the current largely passive approach that relies on assessment of reports filed by device users.

Study context

Population
Donors at 3 clinic sites; batches of Verio test strips from launch of improved-algorithm product to reporting date33 test-strip batches
Participants
506 donors
Outcome
Accuracy per FDA guidance for blood glucose monitoring (BGM) systems, EN ISO 15197:2015 and MARD/MAD
Methodological detailsmanufacturer-led, 3 clinic sites, clinician-led, batches randomly and routinely selected
  • Manufacturer-led
  • 3 clinic sites
  • Clinician-led
  • Batches randomly and routinely selected

Performance

As reported in the abstract

Clinical accuracy vs comparator

FDA: values within ±15% of comparator

98.9%

ISO: within ±15 mg/dL or ±15% at <100 mg/dL (<5.55 mmol/L) or ≥100 mg/dL (≥5.55 mmol/L) glucose, respectively

99.0%

Overall MARD

4.19%

MARD range across all test strip batches

3.54%-5.73%

Original publication

Post-Market Surveillance Assessment of the Clinical Accuracy of a Blood Glucose Monitoring System with an Improved Algorithm for Enhanced Product Performance. J Diabetes Sci Technol. Published 31 August 2021. doi:10.1177/19322968211039465

Composed by AI from the published abstract and checked against it. How Briefs are made →

Get them every week

New studies on this topic in My Week and in your Monday e-mail.

We e-mail you a sign-in link and a code. No password, no spam.