Performance, Safety and Usability of a Third-Generation Continuous Glucose Monitoring System in a Pediatric Population: Results From a Prospective, Multicenter Study

J Diabetes Sci TechnolPublished Prospective single-arm study

Authors' conclusion

In a pediatric population, the novel CGM system demonstrated accurate and stable performance across its 15-day wear life and showed high usability.

Study context

Population
Pediatric participants aged 2 to 17 years, 3 clinical centresper-protocol set
Participants
75 participants
Outcome
MARD vs laboratory venous blood glucose
Methodological detailsmulticentre, per-protocol analysis, Participants randomized to performance assessment on days 1 and 2, days 7 to 9 or days 15 and 16
  • Multicentre
  • Per-protocol analysis
  • Participants randomized to performance assessment on days 1 and 2, days 7 to 9 or days 15 and 16

Main finding

As reported in the abstract

Overall mean absolute relative difference (MARD)

8.89%

MARD days 1 and 2

8.65%

MARD days 7 to 9

8.70%

MARD days 15 and 16

9.30%

Observational data: associations, not effects.

Further findings

DTS error grid: data pairs in zones A+B

Data pairs in clinically acceptable zones A+B
100.0%

True alarm rates

Hypoglycemia
98.6%
Hyperglycemia
98.8%

Sensor wear life

Mean sensor wear life
14.4 days

Usability

High participant/guardian satisfaction with system assembly, sensor insertion and overall comfort

Safety

No device-related or skin-related adverse events were reported

Original publication

Performance, Safety and Usability of a Third-Generation Continuous Glucose Monitoring System in a Pediatric Population: Results From a Prospective, Multicenter Study. J Diabetes Sci Technol. Published 19 July 2026. doi:10.1177/19322968261463536

Composed by AI from the published abstract and checked against it. How Briefs are made →

Get them every week

New studies on this topic in My Week and in your Monday e-mail.

We e-mail you a sign-in link and a code. No password, no spam.