Diabetes technology

Noninvasive Continuous Glucose Monitoring With a Novel Wearable Dial Resonating Sensor: A Clinical Proof-of-Concept Study

J Diabetes Sci TechnolPublished Clinical proof-of-concept study

Authors' conclusion

This proof-of-concept study shows that a novel noninvasive continuous glucose monitor was capable of detecting glucose levels. Furthermore, the ARD results are comparable to first models of commercially available minimally invasive products without the need to insert a needle. The prototype has been further developed and is being tested in subsequent studies.

Study context

Population
Male participants with T1D, age range 19–56 years, at varying levels of glycemiamale, T1D
Participants
29 participants
Comparison
Noninvasive wearable dial resonating sensor vs laboratory venous blood glucose (Super GL Glucose Analyzer)
  • Prototype investigational devicen = 29
Outcome
Co-primary: median and mean absolute relative difference (ARD) across all data points
Methodological detailssingle-arm, open-label, experimental, three stages
  • Single-arm
  • Open-label
  • Experimental
  • Three stages

Main finding

As reported in the abstract

Stage 1 (initial proof-of-principle): median absolute relative difference (ARD)
30%
Stage 1: mean ARD
46%
Stage 2 (improved device design): median ARD
22%
Stage 2: mean ARD
28%
Stage 3 (two consecutive days, no recalibration): median ARD
35%
Stage 3: mean ARD
44%

Original publication

Noninvasive Continuous Glucose Monitoring With a Novel Wearable Dial Resonating Sensor: A Clinical Proof-of-Concept Study. J Diabetes Sci Technol. Published 27 April 2023. doi:10.1177/19322968231170242

Composed by AI from the published abstract and checked against it. How Briefs are made →

Get them every week

New studies on this topic in My Week and in your Monday e-mail.

We e-mail you a sign-in link and a code. No password, no spam.