Reproductive endocrinology

Evaluation of the Siemens Healthineers Atellica TSTII testosterone assay in the CDC HoSt-TT program

J Clin Endocrinol MetabPublished Analytical method evaluation study

Authors' conclusion

The TSTII assay is CDC HoSt-TT certified, has strong alignment to LC-MS/MS, and shows good correlation of calculated FT with equilibrium dialysis/LC-MS/MS. Although LC-MS/MS remains the reference method for testosterone quantification, particularly at low concentrations, CDC HoSt certification of the TSTII assay supports its potential suitability for measuring testosterone concentrations in women and children within defined analytical and clinical contexts.

Study context

Population
HoSt-TT certification samples and remnant patient samples
Comparison
Atellica IM TSTII assay vs ID-LC-MS/MS
Outcome
Compliance of Atellica IM TSTII assay with HoSt-TT criteria and correlation with LC-MS/MS
Methodological detailsmethod comparison, stratified by sex
  • Method comparison
  • Stratified by sex

Main finding

As reported in the abstract

HoSt-TT vs Atellica IM TSTII agreement

Average bias

within ±6.4%

OLS regression slopes

1.02-1.06

Correlation

r ≥ 0.98

Further findings

TSTII analytical performance

Repeatability
2.18% CV
Within-lab precision
3.46% CV
Overall total analytical error
18.3% (IQR 11.6-31.6%)

TT on TSTII vs LC-MS/MS (LDT), native patient samples

Slope
0.9798
Correlation
r = 0.991

Calculated free testosterone (FT) (Vermeulen, ISSAM) vs equilibrium dialysis/LC-MS/MS

Slope
0.9530
Correlation
r = 0.965

Original publication

Evaluation of the Siemens Healthineers Atellica TSTII testosterone assay in the CDC HoSt-TT program. J Clin Endocrinol Metab. Published 19 February 2026. doi:10.1210/clinem/dgag071

Composed by AI from the published abstract and checked against it. How Briefs are made →