Diabetes technology

Continuous Glucose Monitoring in Hospitalized Patients With Type 2 Diabetes: A Step Forward in Inpatient Glycemic Control

Endocr PractPublished Randomised controlled trial

Authors' conclusion

The implementation of CGM in hospitals could represent a significant advancement in diabetes management, offering a more comprehensive and dynamic approach to monitoring glucose levels. Further research is needed to explore the long-term impacts of CGM on clinical outcomes and to optimize its integration into hospital protocols.

Study context

Population
Hospitalized patients with T2D in medical and surgical wards; all used continuous glucose monitoring (CGM) (Abbott FreeStyle2 or FreeStyle3) and 6-point point of care glucose
Participants
37 patients
Comparison
CGM-guided vs point of care glucose-guided glycemic therapeutic decisions
  • An = 19
  • Bn = 18reference
Primary endpoint
Difference in TIR 70-180 mg/dL, hospital glycemia risk index, TBR (<70 and <54 mg/dL) and TAR (>180 and >250 mg/dL)
Methodological detailspilot
  • Pilot

Primary result

As reported in the abstract

TIR 70-180 mg/dL

P = 0.04

Further findings

TAR level 1

P = 0.04

OutcomeABP value
Asymptomatic hypoglycemic events detectedMore detected with Amean 1.65 ± 2.03mean 0.31 ± 0.60P = 0.01

CGM accuracy

Overall mean absolute relative difference
14.7%
Diabetes technology society error grid zone A
69.5%
Diabetes technology society error grid zone B
29.3%

Original publication

Continuous Glucose Monitoring in Hospitalized Patients With Type 2 Diabetes: A Step Forward in Inpatient Glycemic Control. Endocr Pract. Published 13 February 2025. doi:10.1016/j.eprac.2025.02.007

Composed by AI from the published abstract and checked against it. How Briefs are made →

Get them every week

New studies on this topic in My Week and in your Monday e-mail.