Association Between Real-Time Continuous Glucose Monitoring Sensor Active Time and Glycemic Metrics in Patients With Diabetes: A Real-World Setting Analysis
Authors' conclusion
In real-world settings, higher CGM use was associated with better glycemic outcomes, including lower GRI, lower average blood glucose, lower GMI, lower TAR, lower TBR, and higher TIR. Future studies should examine factors influencing sensor use to inform strategies that promote effective CGM adoption.
Study context
- Population
- Patients with diabetes using FreeStyle Libre 2 Plus continuous glucose monitoring (CGM) System, Saudi Arabia, January 2023–September 2024
- Participants
- 12,162 readers
- Exposure
- Sensor usage quintiles
- Outcome
- Trends of glycemic metrics with sensor usage (Jonckheere-Terpstra)
Methodological detailsreal-world, de-identified CGM data
- Real-world
- De-identified CGM data
Main finding
As reported in the abstract
- Glycemia risk index (GRI) decreased with increasing sensor use (Std. J-T)
- −29.672
- Average blood glucose decreased (Std. J-T)
- −23.985
- Glucose management indicator (GMI) decreased (Std. J-T)
- −23.949
- Time above range (TAR) decreased (Std. J-T)
- −21.964
- Time below range (TBR): initial increase then decline from third quantile (Std. J-T)
- −4.903
- Time in range (TIR) increased with greater sensor usage (Std. J-T)
- 25.428
- All P-values
- <.001
Observational data: associations, not effects.
Original publication
Association Between Real-Time Continuous Glucose Monitoring Sensor Active Time and Glycemic Metrics in Patients With Diabetes: A Real-World Setting Analysis. J Diabetes Sci Technol. Published 9 April 2026. doi:10. 1177/ 19322968261427798