Adrenal disease

Adrenocorticotropic Hormone Measured by Immulite 1000 and Elecsys Assay Platforms in Suspected Cases of Hypocortisolism: A Word of Caution

Endocr PractPublished Cross-sectional study

Authors' conclusion

Use of Immulite 1000 platform may result in falsely elevated ACTH values. Physicians should exercise caution while interpreting the results and repeat test using a different assay platform may be considered.

Study context

Population
Patients with haemoglobin E/beta-thalassemia assessed for hypocortisolism; 34 primary, 9 secondary, 33 subclinical adrenal insufficiency (1 μg and 250 μg Synacthen tests)
Participants
104 patients
Exposure
Adrenocorticotropic Hormone (ACTH) measured by Immulite 1000 vs Elecsys Cobas
Outcome
Plasma ACTH
Methodological detailsobservational
  • Observational

Main finding

As reported in the abstract

P < 0.001

Correlation between platforms (r)
0.341, P < .001
Agreement between platforms (Cohen's κ), not strong
0.208, P = .001

Observational data: associations, not effects.

Further findings

Diagnosis of primary adrenal insufficiency

Specificity

Secondary adrenal insufficiency with inappropriately elevated ACTH on Immulite 1000

Patients with secondary adrenal insufficiency with inappropriately elevated ACTH (Immulite 1000)
2

Original publication

Adrenocorticotropic Hormone Measured by Immulite 1000 and Elecsys Assay Platforms in Suspected Cases of Hypocortisolism: A Word of Caution. Endocr Pract. Published 4 March 2025. doi:10.1016/j.eprac.2025.02.018

Composed by AI from the published abstract and checked against it. How Briefs are made →

Get them every week

New studies on this topic in My Week and in your Monday e-mail.

We e-mail you a sign-in link and a code. No password, no spam.