Accuracy and reliability of a continuous glucose monitoring system with a focus on hypoglycaemia

Diabetes Obes MetabPublished Diagnostic accuracy study

Authors' conclusion

CGM accuracy can vary in different clinical hypoglycaemia scenarios. While it remains acceptable in people with diabetes receiving subcutaneous insulin therapy and people with insulinoma, it appears to be inaccurate for glucose monitoring during ITT.

Study context

Population
People with various causes of hypoglycaemia464 continuous glucose monitoring (CGM)/POC glucose pairs
Participants
92 people
Comparison
CGM accuracy across three clinical hypoglycaemia settings: ITT, insulinoma, diabetes on subcutaneous insulin
  • ITTn = 63
  • Insulinoman = 16
  • Diabetesn = 13reference
Outcome
CGM accuracy within the hypoglycaemic range (mean absolute relative difference (MARD), %20/20, diabetes technology society error grid, Bland-Altman analysis)

Performance

As reported in the abstract

CGM accuracy
OutcomeITTDiabetes
MARD50.8%13.9%
%20/2067.5%81.5%
  • ITT vs Diabetes: P < 0.01 — accuracy lower with ITT
  • ITT vs Diabetes: p .02 — accuracy lower with ITT

Further findings

CGM values with moderate risk of failing to detect potentially dangerous hypoglycaemia (error grid) — ITT vs Diabetes · higher with ITT vs diabetes and insulinoma

P < 0.01

OutcomeITTInsulinomaDiabetesP value
Bland-Altman biasITT vs Diabetes · significantly increasing bias with ITT−15.2 ± 13.6 mg/dL— not stated in the abstract−3.3 (10.7) mg/dLP < 0.01

— not stated in the abstract

Original publication

Accuracy and reliability of a continuous glucose monitoring system with a focus on hypoglycaemia. Diabetes Obes Metab. Published 25 December 2025. doi:10.1111/dom.70408

Composed by AI from the published abstract and checked against it. How Briefs are made →

Get them every week

New studies on this topic in My Week and in your Monday e-mail.

We e-mail you a sign-in link and a code. No password, no spam.