Safety and Glycemic Outcomes During the MiniMed TM Advanced Hybrid Closed-Loop System Pivotal Trial in Children and Adolescents with Type 1 Diabetes

Diabetes Technol TherPublished Single-arm trial

Authors' conclusion

In children and adolescents with T1D, MiniMed AHCL system use was safe, A1C was lower, and TIR was increased. The lowest GT and shortest AIT were associated with the highest TIR and lowest TBR and TAR, all of which met consensus-recommended glycemic targets.

Study context

Population
Children and adolescents aged 7–17 years with T1D, 13 investigational centersITT population
Participants
160 children
Comparison
Advanced hybrid closed-loop (AHCL) vs run-in
Outcome
Primary safety and effectiveness endpoints: A1C, TIR, severe hypoglycemia and DKA
Methodological detailspivotal trial, multicentre (13 investigational centers), run-in period
  • Pivotal trial
  • Multicentre (13 investigational centers)
  • Run-in period

Main finding

As reported in the abstract

end of study

A1C reduced from baseline (N = 160) to end of study (N = 136); P < 0.001

7.9 ± 0.9% to 7.4 ± 0.7%

Overall TIR increased from run-in to end of study; P < 0.001

59.4 ± 11.8% to 70.3 ± 6.5%

Further findings

Active insulin time (AIT) 2 h (N = 19) vs longer AIT settings (N = 52), with 100 mg/dL GT

Mean TIR with AIT 2 h
73.4%
TBR <70 mg/dL reduced
3.5% to 2.2%
TAR >180 mg/dL reduced
28.7% to 24.4%

Safety

Severe hypoglycemia and diabetic ketoacidosis (DKA) during AHCL use

Severe hypoglycemia
0
DKA
0

Original publication

Safety and Glycemic Outcomes During the MiniMed TM Advanced Hybrid Closed-Loop System Pivotal Trial in Children and Adolescents with Type 1 Diabetes. Diabetes Technol Ther. Published 25 October 2023. doi:10.1089/dia.2023.0255

Composed by AI from the published abstract and checked against it. How Briefs are made →

Get them every week

New studies on this topic in My Week and in your Monday e-mail.

We e-mail you a sign-in link and a code. No password, no spam.