Invasive validation of the Antares algorithm for determining central blood pressure based on upper arm oscillometric pulse waves in patients with type 2 diabetes
Authors' conclusion
The present study suggests that the Antares algorithm incorporated into the custo screen 400 device can estimate cBP with high accuracy turning a conventional oscillometric BP device into a type II device for the non-invasive estimation of cBP, which is applicable in patients with type 2 diabetes. Integration of Antares into commercially available BP devices could facilitate the introduction of cBP into routine clinical practice as a part of disease and risk management.
Study context
- Population
- Consecutive patients with T2D undergoing elective coronary angiography, November 2017 to September 2020
- Participants
- 119 patients
- Comparison
- Antares-estimated cBP vs invasively measured cBP
- Outcome
- Accuracy of Antares-estimated cBP vs invasively measured cBP
Methodological detailsinvasive validation, simultaneous measurement
- Invasive validation
- Simultaneous measurement
Performance
As reported in the abstract
Mean difference (central systolic BP (cSBP))
1.2 mmHg
±6.3 mmHg
Mean difference (central mean arterial pressure (cMAP))
1.0 mmHg
±4.3 mmHg
Mean difference (central diastolic BP (cDBP))
3.6 mmHg
±5.7 mmHg
- Correlation, cSBP
- r=0.916
- Correlation, cMAP
- r=0.882
- Correlation, cDBP
- r=0.791
- All correlations
- P < 0.001
Original publication