Type 1 diabetes

Automated Insulin Delivery in Elderly with Type 1 Diabetes: A Prespecified Analysis of the Extension Phase

Diabetes Technol TherPublished Prespecified analysis of the extension phase of the AIDE T1D randomised crossover trial

Authors' conclusion

HCL was preferred over PLGS or SAP and remained effective in older adults with T1D.

Study context

Population
Adults ≥65 years with T1D who completed the AIDE T1D crossover trial; mean age 71 ± 4 years (range 65–86), mean T1D duration 42 ± 17 years (range 1–68)of 78 completers
Participants
75 adults
Comparison
Self-selected pump mode
Methodological detailspump mode of participant's choice
  • Pump mode of participant's choice

Primary result

As reported in the abstract

Pump mode selected · additional 3 months

Further findings

TIR 70-180 mg/dL among participants selecting hybrid closed loop (HCL) · RCT and extension phase

TIR similar in the RCT and extension phase
mean 75% ± 10%

Safety

Severe hypoglycaemia

Severe hypoglycemic events reported
1

Original publication

Automated Insulin Delivery in Elderly with Type 1 Diabetes: A Prespecified Analysis of the Extension Phase. Diabetes Technol Ther. Published 11 March 2025. doi:10.1089/dia.2024.0560