Apparent Bilateral Aldosterone Suppression During Adrenal Vein Sampling: An Artefact of the Liaison Chemiluminescence Immunoassay and Other Causes Revealed by Comparisons With Another Commercial Immunoassay and Mass Spectrometry

Clin Endocrinol (Oxf)Published Observational multicentre study

Authors' conclusion

ABAS is a relatively common artefact of the Liaison immunoassay of aldosterone. More rarely, ABAS may reflect sampling blood from an adrenal venous tributary that does not drain from the site of excess aldosterone secretion.

Study context

Population
Patients with primary aldosteronism who underwent bilaterally selective non-stimulated adrenal vein sampling (AVS)
Participants
216 patients
Exposure
Aldosterone by Liaison and iSYS immunoassays vs LC-MS/MS
  • LC-MS/MSn = 216reference
  • Liaisonn = 110
  • iSYSn = 106
Outcome
Apparent bilateral aldosterone suppression (ABAS): aldosterone-to-cortisol ratio below 1.0 in adrenal vs peripheral vein samples, bilaterally
Methodological detailsmethod comparison, bilaterally selective non-stimulated AVS
  • Method comparison
  • Bilaterally selective non-stimulated AVS

Main finding

As reported in the abstract

ABAS rate

P = 0.0004

Observational data: associations, not effects.

Further findings

OutcomeLC-MS/MSLiaisoniSYSP value
ABAS rate, Liaison samplingsLiaison vs LC-MS/MS0.9%14.5%— not stated in the abstractP < 0.0001
ABAS rate, Liaison with LC-MS/MS-measured cortisol substitutedLiaison vs LC-MS/MS0.9%10.9%— not stated in the abstractP = 0.0007
ABAS rate, iSYS samplingsiSYS vs Liaison · iSYS rate a third that of Liaison5.7%— not stated in the abstract4.7%P = 0.0202

— not stated in the abstract

Apparent bilateral aldosterone suppression (ABAS) rate, iSYS samplings

iSYS ABAS rate did not differ from LC-MS/MS

ABAS confined to LC-MS/MS

In one patient resolved by a second sampling revealing pronounced aldosterone secretion vs earlier suppression
1 patient

Original publication

Apparent Bilateral Aldosterone Suppression During Adrenal Vein Sampling: An Artefact of the Liaison Chemiluminescence Immunoassay and Other Causes Revealed by Comparisons With Another Commercial Immunoassay and Mass Spectrometry. Clin Endocrinol (Oxf). Published 29 March 2026. doi:10.1111/cen.70135

Composed by AI from the published abstract and checked against it. How Briefs are made →

Get them every week

New studies on this topic in My Week and in your Monday e-mail.

We e-mail you a sign-in link and a code. No password, no spam.